TRUEresult Meter Specifications
Result Range: 20-600 mg/dL Sample Size: Minimum 0.5 microliter (0.5 μl) Sample: Fresh capillary whole blood, venous blood drawn in EDTA or heparin tubes, or Glucose Control Test Time: Results in as little as 4 seconds Result Value: Plasma values Assay Method: Electrochemical Power Supply: One 3V lithium coin battery #CR2032 (non-rechargeable) Total power when active at full battery = 8.6mW Battery Life: Approximately 2146 tests or 1.5 years Automatic shut-off: After two minutes of non-use Weight: 1.66 ounces Size: 3.44" x 2.16" x 0.69" Memory Size: 500 glucose results
System Operating Range (Meter & Test Strips):
Relative Humidity: 10-90% (Non-condensing) Temperature: 50-104ºF Hematocrit: 20-60%
Note: Use within specified environmental conditions only.
Chemical Composition
TRUEtest Test Strips: Ingredients: Glucose dehydrogenase-PQQ (Acinetobacter calcoaceticus), mediators, buffers and stabilizers.
TRUEtest Glucose Control Solution: Ingredients: Water, d-glucose, buffers, viscosity enhancing agent, salts, dye and preservatives.
|